Call Us Now! 6 012 2151684
For product enquiry or problem? Email us at info@smhealthcare.com.my

PRODUCTS

BENEFUSION VP5 INFUSION PUMP

S-A-S16-VP5

Technical Data

The launch of BeneFusion pump series represents a milestone for Mindray as it moves into the high-end infusion pump market. BeneFusion pumps deliver the answer to user requirements in term of safety, accuracy an intuitive use. The sophisticated design of BeneFusion pump is ready to achieve IT, HIS, CIS connection in the near future and to facilitate one stop service for infusion management for all users.

Physical specification

  • WxHxD        : 245 x 87 x 174 (mm)
  • Net weight    : <2.5kg, Stackable
  • Classification: Type CF, Defibrillation-proof, Class 1

Display 

  • Screen: 3.5" TFT color LCD. 16:9 format
  • Brightness: 1-8 levls adjustable, defailt is level 4
  • Information: Delivery rate, current infusion VTBI, total volume, syringe size, syringe brand, pressure limit, battery capacity, drugs, remaining time, alarms.

CPU type

  • Dual-CPU

Infusion sets

  • Automatic recognition of specified infusion brands or BeneFusion series tubes
  • Applicable to GB 8368-2005 disposals; Diameter 3.5-4.5mm, Thickness 0.8-1.2mm

Basic parameters

  • Mode: 6 modes: Rate mode, Time mode, Body weight mode, Ramp up/down mode, sequentil mode, Micro-infusion mode
  • Flow rate: 0.1-2000 ml/h
  • Increment: 0.01ml(0.1-99,99ml/h); 0.1ml(100-999.9ml); 1ml(1000-2000ml/h)
  • present time: 00:00:01-99:59:59(h:m:s) adjustable
  • Accumulated volume: 0.9999.99 ml
  • KVO: 0.1-5.0ml/h adjustable, step 0.1 ml/h, default is 1.0 ml/h
  • Purge: 0.10-2000ml/h

Function spectrum

  • Automatic calculation of the delivery rate based on different modes
  • Infusion pause/standby: present parameters remain stored
  • Drug library up to 2000 drugs, "ON" and "OFF" selectable; use-defined drug list available
  • History records up to 2000, data could be transmitted to PC and saved as EXCEL format to print out
  • Titration function: change flow rate during infusion with increment 0.01ml/h

Battery:  9 Hours operating Time @25ml/h

Safety system

  • Occlusion 4 pressure unit: mmHg/kPa/bar/psi selectable, auto convertion in different units, default is mmHg, 11 levels occlusion pressure, 150-975 mmHg selectable, default is 450mmHg
  • Auto data lock: "ON" and "OFF", "ON": 1-5min. selectable, step 1 min, default is "OFF",
  • Air detector: 1, volume: minimum 20ul; 2,5 levels  selectable: 50ul,100ul,250ul,500ul,800ul, default is 100ul; 3, Accuracy: 4, alarm: single bubble or accumulated bubble volume in 15 min. reached the sitting limit
  • Anti-bolus: Automaic reduction of th bolus volume following pressure alarm.

Alarm

  • Alarm type: Visual and acoustic alarms, High: occlusion, battery empty, VTBI done, air bubble, door-open, KVO finish, system error, Middle: Standby time expired system abnormal, Low:VTBI near done, Reminder, no battery, low battery, no AC
  • Alarm volume: 1-8 levels selectable, default is level 4
  • Pre-alarms: 1-30min.before infusion is empty, 3min. before battery is completed, 30min. as low battery
  • Reminder: "NO" and "OFF", "NO": 1-5min. selectable, step 1 min.

Interface

  • RS232
  • Nurse call connector
  • DC adapter connector

Power supply

  • 100-240V~, 50/60Hz, 0.53-0.28A; DC Voltage:10-16V, 2.25A-1.5A

Environment requiremt

  • Temperature:5~40oC for operating, -20~60oC for storage
  • Humidity:15~95% for operating, 10~95% for storage
  • Air pressure: 57~106kPa for operating , 50~106kPa for storage
  • Water-proof: IP23 

Lifetime

  • >5 years under continuous duty conditions

Modes specification

  • Flow rate mode: VTBI: 0.1-9999.99ml; Rate: 0.10-2000ml/h
  • Time mode: VTBI: 0.10-9999.99ml; Time: 00:00:01-99:59:59(h:m:s)
  • Body weight mode: Weight: 0.1-300.0kg or 0.1-660.8lb, kg/lb selectable, Drug-Amount: 0.1-999.9,g/mg/ug/ng/lU selectable, default is mg; Volume: 0.10-9999.99ml; Dose: 0.01-999.99; Dose unit: mg,ug.ng,lU/kg/h; mg,ug,ng,IU/kg/min selectable, default is mg/kg/h, VTBI: 0.10-9999.99ml; Time: 00:00:01-99:59:59(h:m:s)
  • Ramp up/down mode: VTBI: 0.10-9999.99ml; Time range: 00:00:01-99:59:59 (h:m:s)
  • Sequential mode: VTBI: 0.10-9999.99ml; Rate: 0.10-1500ml/h; Time: 00:00:01-99:59:59 (h:m:s)
  • Micro-infusion mode: VTBI: 0.10-1000.00ml; Rate: 0.1-100ml/h; Time:00:00:01-99:59:59(h:m:s); Bolus rate: 0.1-100ml/h

 

 

 

Brand: MINDRAY
Packing: 1 UNIT

Additional Info

UNIT