BTL CARDIOPOINT HOLTER H600 SYSTEM
Descriptions:
- ECG lead system (number of leads): 3.7 or 12 leads
- Number of electrodes attached to the patient: 5 or 10 electrodes
- Max. record duration: 1 - 7 days* *12 leads 1 - 4 days
- Certified for use with children lighter than 10 kg: Yes
- Certified for ST segment evaluation: Yes
- Patient activity monitoring: 2 axes accelerometer
- Pacemaker detection: 100 μs / dedicated circuit with 40000 Hz detection function
- Patient button / voice recording: Yes
- Signal quality check: On built-in display or in BTL CardioPoint via USB
- Automatic start of recording: 20 minutes after inserting the batteries
- Automatic continuation of the record after replacing the batteries: Yes
- ECG frequency response: 0.049 - 220 Hz (filters off)
- Sampling frequency: ECG: Delta-Sigma A/D converter, 8 x 2000 Hz Pacemaker: 2 axis detection by dedicated circuit with 40000 Hz function
- Digital resolution: 24 bits / 1.52 μV
- Dynamic range:± 66 mV
- Input signal range: Dynamic range: ±33 mV, polarisation: ±393 mV, consistent voltage offset: ±5 V
- Common mode rejection ratio (CMRR): 5 lead cable: >100 dB, 10 lead cable: >94 dB
- Noise level: < 7 μV
- Dimensions (including patient cable connector): 102 x 62 x 24 mm
- Operating conditions: Temperatuture: 1°C - 55°C, humidity: 10 % - 95 %
- Battery type: 2 x AA > 1.2 V
- Battery capacity / type: Lithium AA: cca 116 hours/Alkaline AA: cca 75 hours/ NIMH 2500 mAh: cca 51 hours
- Ingress protection: IP22
Signal Overview: Graphical user interface
- Fully customizable interface
- Interactive trends
- Preffered lead for analysis setting
- Filter view settings
- Selective view of channels
QRS Morphology:
Evalution tools
- Automatic QRS classification into templates (N, B, V, S, J, artifact)
- Advanced manual QRS labeling
- Template merge
- Group templates re-marking
- ECG phragment view, 3 min interval view, Tiles view
- QRS to template navigation
- Manual measurement with auto snap to P, PQ, QRS, QT
- Graphical split: Yes
- Look for smilar beat: Yes
Rhythm and Arrhytmias:
Evalution tools
- Interactive HR trend
- Detailed data table
- Automatic rhythm detection
- (SR, STachy, SBrady AF, V rhythm, AIVR, VT, SVT, SVER)
- Prematurity beat setting
- Parametrs limit setting
- Interactive patient activity trend: Yes
- Relief map: Yes
- Waterfall: Yes
- Poincare: Yes
- Tachogram: Yes
Ischeme Evalution
- Evalution tools: Interactive ST analysis, ST table, ST deviation trend Parameters limit setting, anual measurements.
- ST level and slope trend per lead: Yes
- ST relief map: Yes
Including:
- Windows 11 Professional c/w i7processor or equivalent specification
- 512GB SSD Hard Drive
- 16GB RAM
- 27'' inch LCD Monitor
- LaserJet Printer
| Code |
Item |
Packing |
Additional Info |
| S-R-B10-H600 |
BTL CARDIOPOINT HOLTER H600 SET (EN, DE, ES, RU), IEC |
1 SET |
|
|
|
| S-R-B10-H600PC |
PC WORKSTATION WITH LASERJET PRINTER |
1 UNIT |
|
|
|
Brand: BTL